Product Cat. No.: GBS-146

For Research Use Only

FOSB(19q13) gene break apart probe reagent Instructions Manual

Product Name

FOSB(19q13) gene break apart probe reagent

Package Specifications

10 Tests/box

Intended Usage

This kit performs fluorescence in situ hybridization staining on the basis of conventional staining, and provides auxiliary information for diagnosis for physicians. The test results are for clinical reference only and should not be used as the sole basis for clinical diagnosis.Clinicians should make comprehensive judgments on the test results based on factors such as the patient’s condition, drug indications,treatment response and other laboratory test indicators.

Detection Principle

The kit is based on fluorescence in situ hybridization technology. A nucleic acid probe islabeled with fluorescein. The target gene is detected with homologous complementary to the nucleic acid probe used. Both after denaturation, annealing and renaturation, the hybrid of the target gene and the nucleic acid probe can be formed, and the qualitative, quantitative or relative positioning analysis of the gene to be measured under the microscope can be performed by the fluorescence detection system.

Product Main Components

The kit consists of FOSB dual-color probes, as shown in Table 1.

Table 1: Kit composition

Component name Specifications Quantity Main components
FOSB dual color probe 100μL/Tube 1 FOSB orange probe ; FOSB green probe

bitmap

Storage conditions & Validity

Keep sealed away from light at -20oC±5oC, and the validity period is 12 months. After the cover is opened, it can be sealed and stored in 2~8°C away from light within 24 hours. After the cover is opened, it should be sealed and stored in -20±5°C away from light for a long time.Transport with temperature below 0°C.

Applicable Instruments

Fluorescence microscopy imaging systems, including fluorescence microscopy and filter sets suitable for DAPI (367/452), Green (495/517),and Orange (547/565).

Sample Requirements

1. Applicable specimen types: Paraffin-embedded specimens from surgical excision or biopsy.
2. The tissue should be fixed with 4% neutral formaldehyde solution within 1 hour after isolation. After tissue fixation, it is routinely dehydrated and embedded in paraffin.

Test Method

1. Hybridization pretreatment
Baking: Slides heating at 80oC for 30min or 65oC for 2h or overnight.
Dewaxing: According to the customer laboratory protocol (Commonly with Xylene for 15min).
Hydration: Take out the slides and put them respectively into 100%, 85% and 70% EtOH at room temperature for 3 minutes each.Take out the slides, and immerse them in deionized water for 3 minutes. Remove the excess of water on the slides by air-drying.
Permeation: Immerse the slides in deionized water at 100oC and boil continuously for 20-40 minutes (Conventional 20min). Remove the excess of water on the slides by air-drying.
Digestion: Protease enzymic digestion at 37°C for 10-40 minutes. Mix the protease work buffer (50mmol HCl) and the 10x protease solution (Pepsin concentration 5%) in a proportion of 9:1 to prepare the enzymatic digestion solution.
Washing: Wash with 2xSSC at room temperature for 5 minutes.
Dehydration: Take out the slides and dehydrate in 70%, 85%, and 100% gradient ethanol at room temperature for 2 minutes each time.Remove the excess of EtOH solution on the slides by air-drying.

2.Denaturing hybridization
The following operations should be performed in a darkroom.

  1. Take out the probe, let it stand at room temperature for 5min, turn it upside down with force, mix the probe well, centrifuge it briefly (do not vibrate with vortex apparatus), drop 10μl into the hybridization area of the cell drop, cover the 22mm×22mm cover glass immediately, the probe should be evenly spread under the cover glass without bubbles, and seal the edge with rubber (the edge sealing must be thorough to prevent the dry slide from affecting the test results in the hybridization process).
  2. Place the glass slide in the hybridization instrument, denature at 85°C for 5 minutes (the hybridizer should be preheated to 85oC) and hybridize at 45°C for 2 to 16 hours.

3. Washing
The following operations should be performed in a darkroom.

  1. Carefully tear off the adhesive around the cover glass with tweezers to avoid sticking off or moving the cover glass.Immerse the cell drop into 2xSSC for about 5s, and take it out. Gently push one corner of the cover glass to the edge of the slide with tweezers, and gently remove the cover glass with tweezers;
  2. The cells were placed at 2xSSC room temperature for 1min;
  3. Take out the slides and immerse in a preheated at 68°C 0.3% NP 40/0.4xSSC (Preparation of 0.3% NP-40/0.4xSSC: For 1L preparation,take 3mL NP 40 and 20mL 20xSSC, dissolve fully, mix well, and use 1M NaOH to adjust the pH to 7.2) solution and wash for 2min.
  4. The sides were immersed in deionized water preheated at 37°C for 1min, and then dried naturally in the dark.

4. Counterstaining
The following operations should be performed in a darkroom
10μL DAPI compound dye is dropped in the hybridization area of the glass slide and immediately covered. The suitable filter is selected for glass slide observation under the fluorescence microscope.

5. FISH results observation
Place the counterstained film under the fluorescence microscope, and first put it under the low-power objective lens (10x) confirm the cell area under the microscope; Go to 40x under the objective lens, find a position where the cells are evenly distributed; Then in the high-power objective (100x) the FISH results of nuclei are observed.

Interpretation of common signal types

FOSB gene site 3 signal
FOSB gene site 3 signal
dots

Negative: 2 fusion

dots

Positive: 1 orange 1 green 1 fusion

Limitations of test methods

1.The results of this kit will be affected by various factors of the sample itself, but also limited by hybridization temperature and time,operating environment and the limitations of current molecular biology technology, which may lead to wrong results.
2.Users must understand the potential errors and accuracy limitations that may exist in the detection process.

Precautions

1. Please read this manual carefully before testing. The testing personnel shall receive professional technical training, and the signal counting personnel must be able to observe and distinguish orange and green signals.
2. When testing clinical samples, when the hybridization signal counting is difficult and the sample is not enough to repeat the retest, or the cell volume is not enough for analysis, the test will not provide the test results.
3. DAPI counterstaining agent used in this experiment has potential toxicity or carcinogenicity, so it is necessary to operate in the fume hood,wear masks and gloves to avoid direct contact.
4. All chemicals are potentially dangerous. Avoid direct contact. Used kits are clinical waste and should be properly disposed off.

Manual Approval date & Revision date

V1. 0: Approval date: April 30, 2020.
V1. 2: Revision date: December 7, 2021.